This paper presents an experimental study for the propagation losses and wave attenuation inside human body for implanted antenna at different body organs. The experimental study discusses the commonly used frequency bands: 402–405 MHz MICS (Medical Implant Communication Service) band, and the industrial, scientific and medical (ISM) band frequency at 915 MHz. The path loss is calculated for different positions of implant at different organs i.e., heart, stomach, and intestine. In addition, the effect of frequency change is discussed and analyzed.
Citation
Huseyin Arslan, Qammer Hussain Abbasi, A. Fatih Demir, Khalid A. Qaraqe, and Erchin Serpedin, "Experimental Characterization of In Vivo Radio Channel at MICS and ISM Bands," Proceedings of 2015 Photonics & Electromagnetics Research Symposium, Prague, July 6 - 9,Page(s)2796-2799
References

1. Beach, R. D., R. W. Conlan, M. C. Godwin, and F. Moussy, "Towards a miniature implantable in vivo telemetry monitoring system dynamically configurable as a potentiostat or galvanostat for two and three electrode biosensors," IEEE Trans. Instr. Meas., Vol. 54, 61-72, 2005.        Google Scholar

2. Bradley, Peter D., "An ultra low power, high performance Medical Implant Communication System (MICS) transceiver for implantable devices," Biomedical Circuits and Systems Conference, IEEE BioCAS, 158-161, November 2006.
doi:10.1109/biocas.2006.4600332        Google Scholar

3. Scanlon, W. G., B. Burns, and N. E. Evans, "Radiowave propagation from a tissue-implanted source at 418 MHz and 916.5 MHz," IEEE Transactions on Biomedical Engineering, Vol. 47, No. 4, 527-534, April 2000.
doi:10.1109/10.828152        Google Scholar

4. Scanlon, W.G., N. E. Evans, and Z. M. McCreesh, "RF performance of a 418-MHz radio telemeter packaged for human vaginal placement," IEEE Transactions on Biomedical Engineering, Vol. 44, No. 5, 427-430, 1997.        Google Scholar

5. Abbasi, Qammer H., Marwa Qaraqe, and Erchin Serpedin, "Numerical characterisation and modeling of in-vivo radio communication," Proceedings of Second Middle East Conference on Biomedical Engineering (MECBME), 365-367, Doha, Qatar, February 2014.
doi:10.1109/mecbme.2014.6783279        Google Scholar

6. Astrin, A. W., H. B. Li, and R. Kohno, "Standardization for body area networks," IEICE Transactions on Communications, Vol. E92-B, No. 2, 366-372, 2009.
doi:10.1587/transcom.e92.b.366        Google Scholar

7. Xia, W., K. Saito, M. Takahash, and K. Ito, "Performance of an implanted cavity slot antenna embedded in the human arm," IEEE Trans. on Antennas and Propagation, Vol. 57, 894-899, 2009.        Google Scholar

8. Gemio, J., J. Parron, and J. Soler, "Human body effects on implantable antennas for ISM bands applications: Models comparison and propagation losses study," Progress In Electromagnetics Research, Vol. 110, 437-452, 2010.
doi:10.2528/pier10102604        Google Scholar

9. Alomainy, Akram and Yang Hao, "Modeling and characterization of biotelemetric radio channel from ingested implants considering organ contents," IEEE Trans. on Antennas and Propagation, Vol. 57, No. 4, 999-1005, 2009.
doi:10.1109/tap.2009.2014531        Google Scholar